- Seen : 212 View
Model
ZH-TSH-120A
Sample limited
The test equipment prohibits the testing of flammable, explosive and volatile substances; the test and storage of corrosive substances, biological tests, strong electromagnetic emission sources etc..
Volume, weight and size
Nominal volume
600 L (other volume models are available)
Inner size
1200×600×500 (mm) W×D×H (other size models are available)
External size
2150×1150×1350 (mm) appr. W×D×H
Weight
340 kg
Working noise
≤ 75db measured in the air, 1 meter from the front of the machine and 1.2 meters from the ground
Power max.
9KW
Current max.
15A
Power supply
· AC380V three-phase four-wire + protective grounding
· Voltage permissible fluctuation range: ± 10% V
· Frequency permissible: (50 ± 0.5) HZ
· TN-S power supply or TT mode power supply
· Protection ground wire grounding resistance is less than 4Ω
· Requires the user to configure the appropriate capacity of the air or power switch for the equipment, and this switch must be independently controlled for the device.
Performance
Functional requirements
1) (air) indoor space heating, humidification, dehumidification, spraying, defogging;
2) To reach the set temperature, humidity, spray in the chamber according to the parameters set;
3) Each function can be operated independently.
Meet the test standards - Test Methods
GB/T20854-2007/ISO14993-2001 Wet and dry salt spray test
GB-T5170.8-2008 Electrical and electronic products environmental test equipment inspection methods - salt spray test equipment
JJB150.11A-2009 - Military equipment laboratory environmental test method - Salt spray test
GBT2424.17-2008 Test Method for Salt Spraying
GB / T 2423.18-2000 Test method for alternating salt spray of sodium chloride
GB-T2423.3-2006 (IEC6008-2-78-2001) Test cab: Constant damp heat test method
GBT2423_4-2008 Test method for heat and moisture alternating test
Temperature and humidity range (climate test)