FEATURE
Three-step defibrillation process
Two-button operation
Extensive voice and visual prompts for the operator
Continuous event recording for reporting each use to a printer or computer
Weekly self-test to ensure readiness
Biphasic energy output
Lock-out protection to prevent inadvertent defibrillation
SPECIFICATION
Defibrillator
Output: Biphasic Truncated exponential
Energy Sequence: 150, 150, 200J
Charge Time: 8 sec. to150J
10 sec. to 200 J(NEW battery)
Analysis Time: 9 sec.
The maximum time from the initation of rhythm analysis to readiness for discharge with a new battery: Less than 30 seconds
The maximum time from the initation of rhythm analysis to readiness for discharge after 6 shocks: Less than 35 seconds
The maximum time from the initation of rhythm analysis to readiness for discharge: Less than 25 seconds
Audible Prompts: 20 audible prompts
Visual Prompts: LED prompts
Controls: Two buttons - On/Off, Shock
Output Energy Accuracy: ±15% into impendance from 25 to 75Ω
The Maximum Voltage: 1200±50V
Output disabled when the patient impedance is outside limits: 20Ω to 200 Ω
Battery
Part No: CR123A-4X2
Non-Rechargeable: 12V DC. 3.0 Ah
Capacity: 100 discharges at 200 Joules or 120 discharges at 150 Joules
Shelf Life (25℃±15℃): 10 years (5 years storage + 5 years standby)
5 years standby (after installation)